14  Specification 
Measuring range: 
0.0 ... 100.0 %O2 (oxygen concentration) 
Accuracy:  
calibrated device at nominal temp.: ± 0.1 %O2  ± 1 digit 
Sensor linearity: < 2 vol.-% +/- 0.1%; < 25 vol.-% +/- 0.5%; 
< 100 Vol.-% +/- 1.0% 
Sensor connection  0.7m connection cable (fixed to device) with jack connector  
Sensor: 
electrochemical partial oxygen pressure sensor, integrated 
in sensor housing (type for replacement: GOEL 370) 
 Response time: 
90% in < 10 sec., depending on temperature 
 Operating life: 
guaranteed 12 months (assuming appropriate usage at air 
pressure) 
 Operating pressure:  0.5 to 2.0 bar absolute. (at one-sided strain: max. 0.25 bar 
over-/under-pressure) 
Meas. Frequency:    about 1 measuring per second 
Display: 
approx. 13 mm high, 3½-digit LCD 
Pushbuttons: 
3 keys for ON/OFF, min-/max-value display, calibration 
Min-/Max- Memory:  min and max measured value are stored 
Nominal temp.: 
25°C 
Ambient temp.: 
0 to 45°C (sensor), -20 to 50°C (device) 
Relative humidity: 
0 to 95 %RH (not condensing) 
Storage 
temperature:
 
-15 to 60°C (sensor), -20 to 70°C (device) 
Power Supply: 
9V-battery type JEC 6F22 (in scope of supply) 
  Power Consumption: 
0.14mA (standard zinc carbon battery: >2100 hours!) 
  Battery Change Indicator:  automatically if battery is used up: "BAT" 
Auto-Power-Off: 
when activated, the device switches automatically off, if it is 
not operated for longer time (selectable 1..120min). 
Housing: 
impact-resistant ABS, transparent panel, front side IP65 
  Dimensions: 
approx. 106 x 67 x 30 mm (L x W x D) without sensor cable 
Weight: 
approx. 185g  incl. battery and sensor 
EMC
  
The device corresponds to the essential protection ratings 
established in the Regulations of the Council for the 
Approximation of Legislation for the member countries 
regarding electromagnetic compatibility (2004/108/EG). 
Device meets EN 61326-1:2006, Additional fault: <1%