Analog input terminals (CH0 In - CH7 In) 
Connect up to eight analog input connections to the screw terminal connections labeled 
CH0 In through CH7 In. Refer to the 
 on page 4-6 for the location of 
these pins.  
You can configure the analog input channels as eight single-ended channels or four 
differential channels. When configured for differential mode, each analog input has 
12-bit resolution. When configured for single-ended mode, each analog input has 11-bit 
resolution, due to restrictions imposed by the A/D converter.  
Single-ended configuration 
When all of the analog input channels are configured for single-ended input mode, eight 
analog channels are available. In single-ended mode, the input signal is referenced to 
signal ground (GND). The input signal is delivered through two wires: 
ƒ
  The wire carrying the signal to be measured connects to CH# IN. 
ƒ
  The second wire connects to GND.  
The input range for single-ended mode is ±10 V, max, with a gain of 2. No other gains 
are supported in single-ended mode. 
Figure 4-4
Figure 4-4. Single-Ended Measurement Connection 
 illustrates a typical single-ended measurement connection. 
GND
CH0
CH1 (differential configuration)
1.5
+
-
 
Single-ended measurements using differential channels 
To perform a single-ended measurement using differential channels, connect the voltage 
to an analog input with an even-number, and ground the associated odd-numbered 
analog input. This configuration is shown in Figure 4-4. 