Functional 
Details 
For example, the miniLAB 1008 exhibits a typical calibrated gain error of ±0.2% on all 
ranges. For the ±10 V range, this would yield 10V × ±0.002 = ±20 mV. This means that 
at full scale, neglecting the effect of offset for the moment, the measurement would be 
within 20 mV of the actual value. Note that gain error is expressed as a ratio. Values 
near ±FS are more affected from an absolute voltage standpoint than are values near 
mid-scale, which see little or no voltage error.  
Combining these two error sources in 
, we have a plot of the error band of 
the miniLAB 1008 for the ±10 V range. This is a graphical version of the typical 
accuracy specification of the product. 
Figure 4-12
Figure 4-12. Error band plot  
The accuracy plots in Figure 4-12 are drawn for clarity and are not drawn to scale 
-FS
+FS
Input Voltage
0
4095
9.77mV
2048
Output Code
Ideal
Ideal +9.77mV + 20 mV
Ideal +9.77mV + 20 mV
Ideal -(9.77mV + 20 mV)
Ideal -(9.77mV + 20 mV)
Ideal
Ideal +9.77mV + 20 mV
 
Channel gain queue 
The miniLAB 1008's channel gain queue feature allows you to set up a scan sequence 
with a unique per-channel gain setting and channel sequence.  
The channel gain queue feature removes the restriction of using an ascending channel 
sequence at a fixed gain. This feature creates a channel list which is written to local 
memory on the miniLAB 1008. This list is made up of a channel number and range 
setting. An example of a four-element list is shown in 
Table 4-4
4-15