Ultrabreitbandantenne UBB27 / Ultrabroadband Antenna UBB27
 
 
© Gigahertz Solutions GmbH, D-90579 Langenzenn  
Revision 1.7 (November 2008)
 Seite 
cant to note which technique was used for 
each measurement. 
 “Rattling tone” for marking of un-pulsed 
transmitters
 
 
When using the HF59B in audio analysis 
mode with the UBB27 attached (The switch 
“Signalanteil” or “Signal” set to “Voll” or 
“Full”), one will almost always hear a rat-
tling tone. This is because sources of un-
pulsed radiation are almost always present 
in the very broad frequency range of the 
UBB27. The loudness of it is proportionate 
to the percentage of un-pulsed radiation in 
the total signal received. The marking is 
done with a frequency of 16 Hz (very low). 
An audio sample can be down-loaded as a 
MP3 file from our home page. 
UBB27 is a perfect match to the Fre-
quency Filter FF6E 
 
It is remotely power supplied by the filter 
through the antenna input, the antenna can 
be assembled to the filter at all times as in 
the “Allpass” setting all its frequency range 
is covered while switchable bandpass fil-
ters allow for an accurate assessment of 
the most important radio services. 
Limits for using the variable frequency 
filters VF2 and VF4 
 
When set to “bypass” the variable fre-
quency filters VF2 and Vf4 have a high pass 
frequency band curve which begins to at-
tenuate frequencies below a couple of 100 
MHz. The analysis in the band of 27 to sev-
eral 100 MHz therefore are to be done 
without VF2 or VF4 mounted to the instru-
ment. 
 
Limits for using the RF amplifiers   
HV10 or HV30 
 
Only the HF59B can supply the Power to 
the UBB27 plus the HV10.  
 
The HV30 cannot be used without an ex-
ternal power supply with the HF59B plus 
the UBB27.  
 
The external attenuator DG20_G3, may be 
used with the UBB27 plus either HFE35B 
or HF59B. 
 
Accuracy 
By itself, the UBB27 inaccuracy range of +/- 
3 dB extends from approx. 85 MHz up to 3.3 
GHz. The antenna continues to work beyond 
that, but with increasing attenuation. 
We state the total accuracy of our HF analyz-
ers for the complete assembly of analyzer 
plus antenna in a far field under well defined 
conditions. (An “average measurement” with 
the complete assembly placed on a non-
conductive support). The measurement inac-
curacies for the complete assembly are the 
following: 
 
HFE35C plus UBB27 is the same +/- 3 dB, 
and 
 
HF59B plus UBB27 increases moderately 
to +/- 4.5 DB. 
 
Below 85 MHz the tolerance level of the 
setup for the calibration becomes predomi-
nant and limits the accuracy achievable for 
the demonstration of the instrument. A simu-
lation, which demonstrated an excellent cor-
relation of actual measurement and simu-
lated signals in the frequency band above the 
lower limit, proves a very good linearity down 
to 27 MHz. Without verification we cannot 
guarantee the accuracy. Frequencies below 
27 MHz are damped out by an internal, ex-
tremely steep, high pass filter 
 
 
 
 
 
 
 
Warranty 
We provide a two-year warranty for factory 
defects on this antenna.  
 
 
 
 
 
 
For 
questions
 and 
service
 please con-
tact for North America: 
 
www.safelivingtechnologies.com 
 
 
For other Countries contact your local  
distributor or: 