Gw Instek PEL-3111 Electronic Load PEL-3111 Hoja De Datos
![Gw Instek](https://files.manualsbrain.com/attachments/dc6af9cceb6bec808125ac96ad99be7dbe37566d/common/fit/150/50/51850356515eec56d4ea258553f26e4c7c86af45115072a82f444014a577/brand_logo.gif)
Los códigos de productos
PEL-3111
PEL-3000 Programming Manual
68
Parameter
HIGH
High range
MIDDLE or MIDD
Middle range
LOW
Low range
Return parameter High
High range
Mid
Middle range
Low
Low range
Example
:CRANge LOW
Sets the current range to Low.
Sets the current range to Low.
Query example
:CRANge?
>Low
The current range is set to Low.
>Low
The current range is set to Low.
[:MODE]:VRANge
Set
Query
Description
Configures the voltage range for all the applicable
operating modes.
operating modes.
Syntax
[:MODE]:VRANge {HIGH | LOW }
Query Syntax
[:MODE]:VRANge?
Parameter
HIGH
High range
LOW
Low range
Return parameter High
High range
Low
Low range
Example
:VRANge LOW
Sets the voltage range to Low.
Sets the voltage range to Low.
Query example
:VRANge?
>LOW
The voltage range is set to Low.
>LOW
The voltage range is set to Low.
[:MODE]:RESPonse
Set
Query
Description
Returns the CV Mode response speed. Fast is the
default response speed.
default response speed.
Syntax
[:MODE]:RESPonse { FAST | SLOW }
Query Syntax
[:MODE]:RESPonse?