Gw Instek PEL-3021 Electronic Load PEL-3021 Hoja De Datos
Los códigos de productos
PEL-3021
COMMNAND OVERVIEW
93
Example
:POWer:L1 MIN
Sets the power value to the minimum.
Sets the power value to the minimum.
Query example
:POWer:L1?
>10W
Returns the “Lewel1” power setting.
>10W
Returns the “Lewel1” power setting.
:POWer:L2
Set
Query
Description
Sets or queries the CP mode “Level2” power. This
command is only applicable to dynamic mode.
Note: A different “Level2” value can be set to
different current ranges.
command is only applicable to dynamic mode.
Note: A different “Level2” value can be set to
different current ranges.
Syntax
:POWer:L2 {<NRf>[ W ] | MINimum | MAXimum }
Query Syntax
:POWer:L2?
Parameter
<NRf>[ W ]
“Level2” power value
MINIMUM or MIN
Minimum power level
MAXIMUM or MAX
Maximum power level
Return parameter <NR2>
Returns the “Level2”
power value.
power value.
Example
:POWer:L2 MIN
Sets the power value to the minimum.
Sets the power value to the minimum.
Query example
:POWer:L2?
>10W
Returns the “Level2” power setting.
>10W
Returns the “Level2” power setting.
:POWer:SET
Set
Query
Description
Sets or queries the CP dynamic mode Set value
when the dynamic units are set to Percent.
This command is only applicable in dynamic
mode.
when the dynamic units are set to Percent.
This command is only applicable in dynamic
mode.
Syntax
:POWer:SET {<NRf>[ W ] | MINimum | MAXimum }
Query Syntax
:POWer:SET?