Gw Instek GSP-9300Spectrum analyzer, spectrum analyzerBandwidth (RBW) 1 Hz - 1 MHz (1-3-10) GSP-9300 Hoja De Datos

Los códigos de productos
GSP-9300
Descargar
Página de 61
 
BASIC OPERATION 
39 
Using the Marker Function 
 
Description 
This section will describe how to activate and 
move a normal marker.  The noise marker 
function will also be used to show how to make 
a basic marker measurement.   
Operation 
1.  Use the procedure described in the previous 
section to display a signal from the REF out 
terminal. 
2.  Press 
Marker
 > Select Marker[F1] and select 
marker number 1. 
3.  Press [F2] and turn the marker 1 on. 
4.  Press Normal[F3] and set the marker position to 
12 MHz using either the keypad, scroll wheel or 
arrow keys. 
5.  Press Function[F5]>Marker Noise[F2] and turn 
the marker noise function on. The noise marker 
function calculates the average noise level over 
a bandwidth of 1Hz, referenced from the 
marker position.