Gw Instek GSP-830Spectrum analyzer, spectrum analyzerBandwidth (RBW) 3/30/300 kHz/4 MHz GSP-830 Hoja De Datos

Los códigos de productos
GSP-830
Descargar
Página de 181
 
 
GSP-830 User Manual 
46
 
Full/Zero Span 
Full or zero span setting sets the span to extreme values: 
3GHz (full) or 0kHz (zero). They provide faster ways to 
view signals in certain situations, such as in time domain 
(zero span) for viewing modulation or in full span for 
viewing signals with unknown frequencies. 
Display full frequency span (3GHz) 
 
Panel operation 
1.  Press the Span key. 
Span
 
 
2.  Press F2 (Full Span). 
Full Span
F   2
 
Range 
3GHz (fixed) 
 
Full span also sets these parameters to fixed values. 
• 
Center frequency: 1.5GHz 
• 
Start frequency: 0.0kHz 
• 
Stop frequency: 3.0GHz 
Display 
Full Span
3.0GHz 
reading
Display zero span (time domain) 
 
Panel operation 
1.  Press the Span key. 
Span
 
 
2.  Press F3 (Zero Span). 
Zero Span
F   3
 
Range 
Center frequency (fixed)