AMPLITUDE
53
1. Press the Amplitude key.
Amplitude
2. Press F5 (Corrections).
Corrections..
F 5
1. Enter correction
edit mode
3. Press F1 (Edit). The display
shows the correction sets.
Edit...
F 1
2. Select correction
set
Press F1 (Select) repeatedly to
select the correction set. 5 sets, 30
points each, are selectable.
select the correction set. 5 sets, 30
points each, are selectable.
Select
1 2 3 4 5
F 1
Example: correction set 3 selected
Select
1 2 3 4 5
F 1
3a. Add correction
point
1. Make sure that the cursor is
pointing to the first empty
frequency point.
frequency point.
Pt. Freq. (MHz) Gain
1
2
3
4
5
2. If necessary, move the cursor
using the Up/Down keys.
3. Enter the frequency using the
Numerical and Unit keys:
9kHz to 3GHz.
9kHz to 3GHz.
BK SP
dB
Enter
GHz /
Sec
mSec
MHz /
kHz /
Sec
μ
4. The cursor automatically
moves to the Gain side.
Enter the gain using the
Numerical keys and dB key
Range: −40dB to +40dB.
Enter the gain using the
Numerical keys and dB key
Range: −40dB to +40dB.
Pt. Freq. (MHz) Gain
1
2
3
4
5
2200
BK SP
dB
Enter
GHz /
Sec
mSec
MHz /
kHz /
Sec
μ
5. Repeat the above procedure for all correction data.
The points are automatically sorted by frequency
(low → high).
(low → high).