   AMPLITUDE 
   53
 
1.  Press the Amplitude key. 
Amplitude
 
2.  Press F5 (Corrections). 
Corrections..
F   5
 
1. Enter correction 
edit mode 
3.  Press F1 (Edit). The display 
shows the correction sets. 
Edit...
F   1
 
 
 
2. Select correction 
set 
Press F1 (Select) repeatedly to 
select the correction set. 5 sets, 30 
points each, are selectable. 
Select
1  2  3  4  5
F   1
 
Example: correction set 3 selected 
 
Select
1  2  3  4  5
F   1
 
3a. Add correction 
point 
1.  Make sure that the cursor is 
pointing to the first empty 
frequency point. 
Pt. Freq. (MHz) Gain
1
2
3
4
5
 
 
2.  If  necessary, move the cursor 
using the Up/Down keys. 
 
 
3.  Enter the frequency using the 
Numerical and Unit keys: 
9kHz to 3GHz. 
BK  SP
dB
Enter
GHz /
Sec
mSec
MHz /
kHz /
   Sec
μ
 
 
4.  The cursor automatically 
moves to the Gain side. 
Enter the gain using the 
Numerical keys and dB key 
Range: −40dB to +40dB. 
Pt. Freq. (MHz) Gain
1
2
3
4
5
2200
 
BK  SP
dB
Enter
GHz /
Sec
mSec
MHz /
kHz /
   Sec
μ
 
 
5.  Repeat the above procedure for all correction data. 
The points are automatically sorted by frequency 
(low  → high).