Gw Instek GSP-830Spectrum analyzer, spectrum analyzerBandwidth (RBW) 3/30/300 kHz/4 MHz GSP-830 Hoja De Datos
Los códigos de productos
GSP-830
FREQUENCY/SPAN
43
Display
Center
Frequency
reading
reading
Set frequency span
Panel operation
1. Press the Span key.
Span
2. Press F1 (Span).
Span
3 GHz
F 1
3. Enter the value using the
Numerical and Unit keys, Arrow
keys, and Scroll knob.
keys, and Scroll knob.
BK SP
dB
Enter
GHz /
Sec
mSec
MHz /
kHz /
Sec
μ
Range
2kHz to 3GHz
* Arrow keys & Scroll knob resolution: 1-2-5 sequence
(0 [zero span], 2kHz, 5kHz, 10kHz, 20kHz, 50kHz,
.....1GHz, 2GHz, 3GHz)
* Arrow keys & Scroll knob resolution: 1-2-5 sequence
(0 [zero span], 2kHz, 5kHz, 10kHz, 20kHz, 50kHz,
.....1GHz, 2GHz, 3GHz)
Display
Span
Span
reading
reading
Note
•
Center frequency and span settings automatically
change according to start and stop frequency settings,
and vice versa.
change according to start and stop frequency settings,
and vice versa.
•
If the span becomes smaller than the CHBW (channel
bandwidth) in the ACPR or OCBW measurement
(page82), the warning “Span is less than CHBW!”
appears in the command window.
bandwidth) in the ACPR or OCBW measurement
(page82), the warning “Span is less than CHBW!”
appears in the command window.