 Note 
 
The load can be set to automatically turn on at 
start up. See page 75. 
 
The load can be turned on via remote control. 
See the programming manual. 
 
 The load can be turned on via external control. 
See page 143. 
 
By default the load will automatically turn off if 
the range or operating mode (CC, CV, CR, CP) is 
changed. To disable this behavior, Set Load Off 
(Mode)
 and Load Off (Range) to the OFF setting. 
See page 76 for details. 
Display 
Mode
CP
IRange
H 35A
VRange
L 15V
Function
Dynamic
Configure
1.500
V
0.000
A
1.50
w
Timer1
0.025
Level2
0.00
Level1
0.00
LOAD
RS232
Fine
03/Sep/2012
mS
W
W
LOAD on
 
Shorting the Load 
 
Description 
The Short key can be used to simulate a short 
circuit of the load input terminals. A short 
circuit is simulated by: 
 
Setting the current to the maximum value in CC 
mode. 
 
Setting the resistance to the minimum value in 
CR mode. 
 
Setting the voltage to the minimum value in CV 
mode. 
 
Setting the power to the maximum value in CP 
mode.    
 
When the load is shorted, the external controller 
also sends a short signal. See page 149 for usage 
details.