Gw Instek PEL-3021 Electronic Load PEL-3021 Hoja De Datos
Los códigos de productos
PEL-3021
PEL-3000 User Manual
68
Display
Mode
CC
I Range
H 35A
V Range
H 150V
Function
Dynamic
Configure
0.00
V
0.000
A
0.00
w
Timer1
0.025
SlewRate
2500.00
SlewRate
2500.00
LOAD
RS232
03/Sep/2012
ms
mA/uS
mA/uS
Slewrate settings
Slewrate settings
CV Mode Response Speed
Description
The response speed setting is the response
speed for the negative feedback control of the
load current when used in CV mode. Response
speed settings are only applicable to CV mode.
speed for the negative feedback control of the
load current when used in CV mode. Response
speed settings are only applicable to CV mode.
A response speed that is too fast could cause the
unit to be unstable.
unit to be unstable.
Reducing the response speed can improve
stability.
stability.
Operation
1. Make sure the load is off.
2. Press
Main
. Make sure the unit is in CV
mode by using the Mode[F1] soft-key.
3. Select the response speed with the Response[F4]
soft-key.
Range:
Slow, Fast