Gw Instek PEL-3021 Electronic Load PEL-3021 Hoja De Datos
Los códigos de productos
PEL-3021
PEL-3000 User Manual
74
Operation
1. Press
Main
> Configure[F5] > Other[F2] and
turn the Count Time on or off.
Range:
ON, OFF
Display
Mode
CP
IRange
H 35A
VRange
L 15V
Function
Dynamic
Configure
0.000
V
0.000
A
0.00
w
Timer1
0.025
Level2
0.00
Level1
0.00
LOAD
RS232
Fine
03/Sep/2012
mS
W
W
0:00:05
Elapsed time
Cut Off Time
Description
The Cut Off Time function will turn the load off
after a set-amount of time. After the load has
been turned off, a popup screen will display the
voltage level when the load was turned off.
after a set-amount of time. After the load has
been turned off, a popup screen will display the
voltage level when the load was turned off.
Operation
1. Press
Main
> Configure[F5] > Other[F2] and
set the Cut Off Time.
Range:
OFF, 1 second ~ 999 hours:59
minutes:59 seconds
minutes:59 seconds
Display
Mode
CP
I Range
H 35A
V Range
L 15V
Function
Dynamic
Configure
0.000
V
0.000
A
0.00
w
Timer1
0.025
Level2
0.00
Level1
0.00
LOAD
RS232
Fine
03/Sep/2012
mS
W
W
0:00:05
Cut off time
Time Up
Voltage : 5.1223V
Enter
Voltage at cut off time