Kern Counting scales CPB 15K0.2N Weight range 15 kg Readability 0.2 g mains-powered, rechargeable Silver CPB 15K0.2N Manual De Usuario
![Kern](https://files.manualsbrain.com/attachments/04ab1b8985747ba02c8b0a2770613923db7d4965/common/fit/150/50/4233fd5bdc88993dc151edc4fd729c0efa33c251fd07d340e43459f4e2b9/brand_logo.gif)
Los códigos de productos
CPB 15K0.2N
CPB-N / CPB-DM-BA-e-1323
20
Press TARE button, the acceleration
value due to gravity is displayed.
Press the TARE key anew
„LoAd 0“ appears, after a short time a signal
will sound and „LoAd 1“ appears
will sound and „LoAd 1“ appears
First adjustment weight (s. tab. 1)
after a short time a signal will sound and
„LoAd 2“ appears
„LoAd 2“ appears
Second adjustment weight
after a short time again a signal will sound
and „LoAd 3“ appears.
and „LoAd 3“ appears.
Third adjustment weight
after a short time again a signal will sound
and „LoAd 4“ appears.
and „LoAd 4“ appears.
Forth adjustment weight
after a short time again a signal will sound
and „LoAd 3“ appears.
and „LoAd 3“ appears.
Forth adjustment weight
after a short time a signal will sound and
„LoAd 2“ appears.
„LoAd 2“ appears.
Third adjustment weight
after a short time a signal will sound and
„LoAd 1“ appears
„LoAd 1“ appears
(Example)