Greisinger GMH3691 and GGO369 Oxygen Meter with Sensor 601667 Hoja De Datos
Los códigos de productos
601667
31
Handheld instrument
Display / Controller
Logger / EASYBus
Transmitter
Temperature probe
Alarm / Protection
Residual oxygen meas. device
for quick and cost-effective measure-
ment of residual oxygen
GMH
3691
GOG
Application:
Essentially there, where delicate products are
conserved by low-oxygen atmospheres (protec-
tive gas), this instrument is suitable to check the
residual oxygen content.
packaging industry
food industry
Specifi cation:
(summary)
Meas. range: 0,0 ... 100,0 % O
2
(O
2
-concentration)
Accuracy: (whole system - during carefully
calibration and measuring)
1-point-calibration: ±0.2 %O
2
±1 digit
(for
concentrations
<
10%)
2-point-calibration: ±0.1 %O
2
±1 digit
(for
concentrations
<
10%)
Oxygen probe: Oxygen-partial pressure probe,
built in external sensor housing
Response time: t
90
< 10 sec., depending on
temperature
Operation life:
warranty for sensor element
12 months
(appropriate appli-
cation and ambient pressure)
Working pressure: 0.5 to 2.0 bar abs.
Over-/under-pressure: max. 0,25 bar
Working temperature: 0 to 50°C (sensor),
-20 to 50°C (device)
Relative humidity: 0 to +95%RH
(non-condensing)
Storage temperature: -15 to 60°C (sensor),
-20 to 70°C (device)
Power supply: 9V battery type IEC 6F22
Dimensions case: approx. 394 x 294 x 106 mm
Weight: approx. 1400g (cpl. set)
for additional technical data refer to
GMH3691 and GGO369
Scope of supply:
Instrument GMH3691, hand pump with air tube,
GOG oxygen sensor with penetration needle, case
GKK3500, spare needle ø0,9mm, rubber foam
Spare elements, accessories:
GOG-SET
Set without instrument
Scope of supply: GOG oxygen sensor with
penetration needle, hand pump with air tube, case
GKK3500, spare needle and 40 rubber foam sticker
GOEL 369
spare sensor element
GOG-N
needle,
Ø
0.9 mm (5 pieces)
GOG-A
rubber foam sticker (40 pieces)
ST-R1
device protection bag
with cut-out for probe connection
for add. accessories p.r.t. page 40/41
Atmospheric oxygen sensores
for devices of the GMH369x series
Spare elements, accessories:
GOEL 369
spare sensor element for replacement by user
GOEL 370
spare sensor element for replacement by user
GOEL 369 S
spare sensor element for replacement by user
ESA 369
spare tube-adapter M16x1, for tubes with a inner-diameter of 15mm
GOO 369
for universal application
GOO 370
sensor for diving application
GOO 369 S
O
2
sensor for high CO
2
concentration
closed sensor type
open sensor type
suitable for under and over pressure
for using in gas-tight systems
Application:
Suitable for measuring in normal atmosphere
and in systems without or with slight under or
over pressure. The sensor type features a screw
thread and can be built in gas-tight in almost every
system directly resp. with tube-adapter
suitable for air- or gas-stream
quick temperature compensation
Application:
Because of the special sensor construction the
measuring gas streams optimally around the sen-
sor and escapes through holes in the housing into
the air. No pressure build-up at slight streaming of
the probe, that falsify the result of measurement.
Particularly suitable for measuring of gas out of
gas-bottle etc. Even measuring indoor-gas con-
centration is possible.
GGO 369
for universal application
GGO 370
sensor for diving application
GGO 369 S
O
2
sensor for high CO
2
concentration
Options:
(for all types)
cable length 4m
upcharges
cable length 10m
upcharges
Specifi cation:
GGO/GOO 369
GGO/GOO 370
GGO/GOO 369 S
Application:
standard
diving
CO
2
containing gases
Specifi c features:
Stronger membrane
Acidic electrolyte
Coated electronics
independence on operating position
temperature compensation
Measuring range:
Partial oxygen pressure: 0 ... 1100 hPa O
2
0 ... 1100 hPa O
2
0 ... 300 hPa O
2
Oxygen concentration:
0,0 ... 100,0 % O
2
0,0 ... 100,0 % O
2
0,0 ... 25,0 % O
2
Temperature:
0,0 ... 50,0 °C
0,0 ... 45,0 °C
0,0 ... 50,0 °C
Response time: t
90
<5 sec.
<10 sec.
<15 sec.
Operating conditions:
0 - 50 °C
0 - 45 °C
0 - 50 °C
0 - 95 %RH
0 - 95 %RH
0 - 95 %RH
Ambient pressure:
0,5 to 2,0 bar abs.
0,5 to 2,0 bar abs.
0,5 to 2,0 bar abs.
Over-/under-pressure: max.
0,25 bar
(pressure difference sensor membrane to ambient – sensor screwed-in)
Storage temperature:
-15 to +60 °C
Operation life:
approx. two years (warranty for sensor element: 12 months)
Sensor:
GOEL 369
GOEL 370
GOEL 369 S
Oxygen-partial pressure probe, mounted in external sensor housing
Connection:
approx. 1,3 m cable with Mini-DIN-plug.
Dimensions of housing:
approx. Ø 36 mm Housing with M16 x 1-screw thread
(sensor can be connected to line tubes by means of an additional adapter)
length: approx. 91 mm (141 mm incl. anti-buckling glanding)
Weight:
approx. 135 g
Scope of supply:
sensor, tube-adapter, fl ow diverter
N E W
N E W