1.6.  Features and Benefits 
 
• 
Powered by 4 ‘AA’ alkaline batteries or NiMH rechargeable batteries. An 
optional AC adapter provides extended operation. 
• 
O
2
  and CO
LOW
  measurement standard. Optional measurement of up to two 
additional gases, including CO
HIGH
, NO, NO
2
, or SO
2
• 
With the appropriate sensors installed, the analyzer optionally displays 
pollution conversions for CO, NO, NO
2
, and SO
2
.  Pollution conversions 
include ppm, #/MBTU, mg/m
3
, and g/GJ. 
• 
B-Smart
®
 sensor technology allows pre-calibrated sensors to be installed in 
the field.  Sensors are provided with data that can be entered through the 
PCA
®
3 software, or instrument calibration menus, for easy calibration. 
• 
Automatic flushing of the CO
LOW
  sensor with fresh air if the CO level 
exceeds 4,000 ppm, thus protecting the CO
LOW
 sensor from high CO levels. 
To measure CO levels above 4,000 ppm, the analyzer automatically 
switches to its CO
HIGH
 sensor, if installed. 
• 
Automatic purging of the gas-sample system if the detected gas levels are 
abnormally high when the analyzer is turned OFF. 
• 
Displays temperatures in either °F or °C 
• 
Displays pressure in either inwc, mb, Pa, or hPa 
• 
Backlit color graphic LCD with zoom capabilities 
• 
Low battery alarm 
• 
Stores 500 individual combustion records, which can later be recalled for 
viewing, printing, or downloading to a personal computer. Stores an 
additional 500 data logged records. 
• 
Wireless IrDA link for printing  current and stored combustion records, 
pressure records, calibration data, and diagnostic data 
• 
USB connectivity for downloading data to personal computer 
• 
Field replaceable sensors and thermocouple 
• 
Two year warranty on analyzer and all gas sensors except the O
2
  sensor 
which has a one (1) year warranty. 
• 
Language options including English, French, and Spanish 
• 
Custom Display Formats 
• 
Calibration Reminders: PCA
®
3 can be configured to remind the user that 
calibration is past due. 
• 
Auto/manual zeroing option allows the operator to select the start-up mode.  
Auto-zero determines the zero reading of the CO channel in fresh air and 
provides an offset for the CO measurements.  Also, all sensing channels are 
automatically zeroed on ambient air when the analyzer is first turned  ON.  
Manual zero is used to detect CO that may be present during start-up.