Bacharach PCA 3 Manual De Propietario

Descargar
Página de 100
PCA
3 Manual 
 
 
0024-9472 Rev 2 
 
47 
5.8.  Performing a Draft / Pressure Measurement 
 
The difference in pressure (
∆P) between two areas can be measured by using 
the analyzer’s two pressure ports and the PRESSURE screen. By using the –
∆P 
port as the reference, the pressure applied to the +
∆P port will be displayed on 
the PRESSURE screen as the differential pressure between the two ports.  
Perform a draft / pressure measurement as follows. 
1.  Turn ON the analyzer and allow it to complete its warm-up cycle (see 
 
NOTE:  Refer to Pressure Units Selection  on  page  26  for 
information on selecting the pressure units of measure. 
2.  Display the MAIN MENU by pressing the MENU (F2) button. If 
necessary, press ESC until MENU appears above F2.  
3.  Use the 
▲▼ buttons to highlight PRESSURE, and then press ENT to 
display the PRESSURE screen. 
4.  Before taking a measurement, the pressure sensor may need to be re-
zeroed if it is not already displaying zero with both pressure ports open 
to the atmosphere. If necessary, zero the pressure sensor as follows. 
• 
Press the 
ZERO (F2)
 button. 
• 
Disconnect any hoses connected to the +
∆P and –∆P 
ports, and then press 
ENT
 to zero the pressure sensor. 
• 
Reconnect any hoses. When measuring draft, simply 
leave the -
∆P port open to the atmosphere and connect 
the probe’s draft hose to the +
∆P port (see Figure 5-12). 
 
   
   
 
Figure 5-10.  Performing a Draft/Pressure Measurement
 
5.  Do one of the following to measure draft or differential pressure: 
• 
To measure draft, simply insert the probe into the stack 
and observe the draft reading on the PRESSURE screen. 
• 
To measure differential pressure, connect two sampling 
hoses to the +
∆P and –∆P ports, and place the open end 
of each hose into the areas being measured. The