0024-9472 Rev 2 
 
Figure 5-14.  Emptying the Water Trap
 
 
5.12.  Data Logging 
 
5.12.1.  Overview 
When the logging function is activated, up to 500 combustion test records will be 
automatically stored in memory at a preset interval over a predetermined 
duration.  
The maximum duration that data can be collected is determined by the interval. 
For example, if the interval is set to 10 seconds, then the maximum selectable 
length of time in which data can be collected to fill 500 memory locations would 
be 1
hour (500 x 10 seconds = 5000 seconds or 83
minutes). If the operator 
chooses a duration that is longer than possible for the interval chosen, then the 
analyzer automatically selects the highest duration possible for the selected 
interval. 
The stored logged data can either be viewed on the display using the analyzer’s 
memory function (see  Memory Functions  on  page  54), or downloaded to a 
personal computer using the supplied Fyrite
®
  User Software and USB cable. 
Downloaded data is stored on the computer’s hard drive as a comma separated 
value ASCII text file with a CSV extension, which can be opened by most 
spreadsheet programs for analysis.