5.12.4.  Starting the Data Logging Process 
 
After turning ON data logging and setting the interval and duration, the analyzer 
will pause at the LOGGING SUMMARY screen, where the currently selected 
interval and duration time periods are displayed. 
Press 
ENT
  to start the combustion test and logging process. At this time the 
Combustion Test LOG screen will appear, indicating that the analyzer is now 
performing a combustion test and the data is being stored in memory.  
 
 
NOTE:  At the top of the screen, the memory location where the 
current log entry is being saved is displayed.  
 
 
NOTE:  If there were previously stored log entries, each new logging 
session will start to store data in the next available memory location. 
 
 
NOTE:  At no time will new data overwrite old data.  Logging will 
stop when available memory is filled. 
 
    
 
Figure 5-17.   Starting the Data Logging Process
 
5.12.5.  Ending the Data Logging Process 
 
Data logging will stop and the pump will turn OFF after the prescribed duration, 
or after all 500 memory locations are filled.  
 
To exit the Combustion Test LOG screen, press the 
RUN/HOLD
 button twice to 
display the Combustion Test HOLD screen. 
 
 
NOTE:  To end the logging process at any  time, press the ESC 
button.