Bacharach PCA 3 Manual De Propietario
PCA
3 Manual
56
0024-9472 Rev 2
Figure 5-19. Recalling Logged Test Data
5.13.4. Clearing Memory
When all memory locations used to store individual combustion test records have
been filled, the next combustion test record saved will not overwrite the oldest.
When all logging memory locations in the logging directory are full, they must be
manually cleared in order to store new data. At no time will the logging process
overwrite older data. Logging will stop when available memory is filled. Each
memory bank can be individually cleared, or all memory locations in both banks
can be cleared simultaneously. Do the following to clear memory:
been filled, the next combustion test record saved will not overwrite the oldest.
When all logging memory locations in the logging directory are full, they must be
manually cleared in order to store new data. At no time will the logging process
overwrite older data. Logging will stop when available memory is filled. Each
memory bank can be individually cleared, or all memory locations in both banks
can be cleared simultaneously. Do the following to clear memory:
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the
▲▼ buttons to highlight MEMORY, and then press ENT to
display the MEMORY MENU.
Figure 5-20. Clearing Memory
3. Use the
▲▼ buttons to highlight one of the following:
•
CLEAR MEMORY DATA clears only the individual saved
combustion test records.
combustion test records.
•
CLEAR LOGGING DATA clears only the combustion test records
that were saved during the logging process.
that were saved during the logging process.
•
CLEAR ALL MEMORY clears all memory locations in both memory
banks.
banks.
4. Press ENT to display the CLEAR MEMORY, LOGGING, or ALL DATA
conformation screen. Highlight YES to confirm that memory is to be
cleared, and then press ENT to actually clear memory as evidenced by
the display of the “WAIT Erasing memory” screen.
cleared, and then press ENT to actually clear memory as evidenced by
the display of the “WAIT Erasing memory” screen.