Bacharach PCA 3 Manual De Propietario

Descargar
Página de 100
PCA
3 Manual 
 
 
56 
 
0024-9472 Rev 2 
   
   
 
Figure 5-19.   Recalling Logged Test Data
 
5.13.4.  Clearing Memory 
 
When all memory locations used to store individual combustion test records have 
been filled, the next combustion test record saved will not overwrite the oldest.  
When all logging memory locations in the logging directory are full, they must be 
manually cleared in order to store new data. At no time will the logging process 
overwrite older data.  Logging will stop when available memory is filled.  Each 
memory bank can be individually cleared, or all memory locations in both banks 
can be cleared simultaneously.  Do the following to clear memory: 
1.  Display the MAIN MENU by pressing the MENU (F2) button. If 
necessary, press ESC until MENU appears above F2
2.  Use the 
▲▼ buttons to highlight MEMORY, and then press ENT to 
display the MEMORY MENU. 
   
   
 
Figure 5-20.   Clearing Memory
 
3.  Use the 
▲▼ buttons to highlight one of the following: 
• 
CLEAR MEMORY DATA clears only the individual saved 
combustion test records. 
• 
CLEAR LOGGING DATA clears only the combustion test records 
that were saved during the logging process. 
• 
CLEAR ALL MEMORY clears all memory locations in both memory 
banks.  
4.  Press ENT to display the CLEAR MEMORY, LOGGING, or ALL DATA 
conformation screen. Highlight YES to confirm that memory is to be 
cleared, and then press ENT to actually clear memory as evidenced by 
the display of the “WAIT Erasing memory” screen.