6.11.  CO
HIGH
 Sensor Calibration 
 
This procedure spans the optional carbon monoxide high sensor (4,001 to 
20,000
 ppm) to a known gas level. 
6.11.1.  Material Required 
• 
Calibration Kit 
(P/N 0024-7059) 
• 
Gas Cylinder 
(4,000 ppm CO in Air, P/N 0024-1155) 
 
 
Figure 6-12.   Gas Sensor Calibration Equipment
 
 
6.11.2.  Procedure 
1.  If not already done, turn ON the analyzer and display the 
CALIBRATION LIST screen (refer to Calibration on page 62). 
2.  Use the 
▲▼ buttons to highlight CO-HI, and then press ENT to display 
the CALIBRATE CO-HI screen. 
 
NOTE:  “Measured” is the current CO reading, while 
“Applied” is a known CO level that will be applied for 
calibration purposes. 
3.  Attach a 4,000
 ppm CO cylinder to the regulator of the calibration fixture 
4.  Use the 
▲▼ and ◄► buttons to enter an applied value that exactly 
equals the concentration on the CO cylinder.