7.5.  Sensor Replacement 
 
All sensors are replaced in a similar manner. Do the following to replace either 
the O
2
, CO
LOW
, CO
HIGH
, NO, NO
2
, or SO
2
 sensor. Refer to Replacement Parts on 
page 90 for a list of sensors and part numbers. 
 
 
NOTE:  The typical life expectancy of the O
2
, SO
2
 and NO
2
 sensors 
is 2 years.  Typical life expectancy of the CO, CO
HIGH
  and NO 
sensors is 3 years. 
 
1.  Turn OFF the analyzer. 
2.  Remove the battery cover. 
3.  Remove the sensor retainer. 
4.  Remove the tubing from the gas cup of sensor being replaced. 
5.  Unplug the sensor along with its printed circuit board from the analyzer’s 
main board. 
6.  Remove the sensor’s gas cup by twisting its bayonet fitting 
counterclockwise, and then pulling the cup straight up. 
7.  Carefully remove the old sensor from the B-Smart® sensor PCB, 
keeping in mind that there is an adhesive foam disk in between the PCB 
and sensor. 
 
 
IMPORTANT:  When replacing the O
2
  sensor, be sure that the “+” 
sensor pin plugs into the PCB socket that is also marked “O
2
+”.  
Also, when replacing an NO sensor, replace the bias battery. 
 
8.  Remove the replacement sensor from its plastic canister. Note that the 
CO
LOW
, CO
HIGH
, SO
2
, and NO
2
 sensors have a shorting spring installed 
between two of their pins. Remove and discard this spring before 
installing the sensor. 
9.  Attach the gas cup to new sensor, making sure that when the sensor 
assembly is plugged into the analyzer’s main board, the tubing 
connections on top of gas cup are aligned as shown in Figure 7-7. 
10.  Plug the sensor into analyzer, and then reattach tubing to gas cup. 
11.  Install the sensor retainer and battery cover.