CHAPTER 8.  TROUBLESHOOTING 
 
8.1.  Error Symbols 
 
Error symbols are shown in the data fields of the Combustion Test screens and 
on the printout to indicate the following. 
 
Error Symbol 
Description 
-   -   - 
Three hyphens or dashes indicates that the calculated data cannot 
be displayed because the measured data necessary to make the 
calculation is out of range. For calculations to occur, the oxygen level 
must be below 16% and the stack temperature must be below 
2,000
 °F (1,093 °C). 
* * * 
Three stars appear in the data field of sensors that are not installed, 
defective, or found to be in error during warm-up, and also in the data 
fields of that sensor’s related calculated values. For example, if the 
data field of either the NO or NO
2
 sensor contains three stars, then 
the NOx data field will also contain three stars. 
X   X   X 
Indicates sensor over range. 
 
8.2.  Diagnostics and Status Screens 
 
The DIAGNOSTICS menu provides information regarding the operation of the 
analyzer.  Information includes the following. 
 
Diagnostic 
Description 
Time Meters 
Displays the run time of the 
analyzer, sample pump and 
purge pump in hours of 
operation. 
 
 
Figure 8-1.   Time Meters
 
Main 
Diagnostics 
Lists the current status of the Stack and Air Thermocouple channels, 
Reference Temperature channel, Pressure channel, and Battery. 
O
2
 Sensor Life 
Displays the approximate remaining life of the Oxygen sensor. 
B-Smart 
Sensors 
Displays the current status of all B-Smart sensors installed. 
Fresh Air 
Diagnostics 
Fresh air diagnostics will cycle the instrument through the 60 second 
warm-up to check on the status of all sensors.  The probe must be in 
fresh air to perform a valid check.