Series 830/840/860 Instruction Manual 
Appendix A
 
IM-83/84/86-H 
A-3
 
Calculating For a Single Gas 
The following tables provide K-factors and thermodynamic prop-
erties of gases commonly used with mass flow meters and control-
lers.  The purpose of these tables is two-fold: 
 
1.  Calibrating an “actual” gas with a reference gas. This is par-
ticularly useful if the actual gas is not a common gas or if it is 
toxic, flammable, corrosive, etc. 
 
2.  Interpreting the reading of a flow meter or flow controller 
which has been calibrated with a gas other than the actual gas. 
 
In applying the tables, the following fundamental relationship is used: 
 
 
 
 
Q
1
/Q
2
 = K
1
/K
2
  
 
 
     (1) 
 
Where: 
 
Q  =  The volumetric flow rate of the gas referenced to standard 
 
conditions of 0°C and 760 mm Hg (sccm or slm), 
 
K  =  The K-factor defined in equation (6), 
 
( )
 1
  =  Refers to the “actual” gas, and 
 
( )
 2
  =  Refers to the “reference” gas. 
 
The K-factor is derived from the first law of thermodynamics ap-
plied to the sensor tube, as described in Chapter 1: 
 
H  =
mC  !T
N
p
(2)