Anritsu ME7840A Manual De Usuario

Descargar
Página de 126
5-6
S-PARAMETER
TEST: K FACTOR
This test describes the K Factors measurement. This measurement is
described in Chapter 1, paragraph 1-10.
Procedure.
Prepare the ME7840A as shown in Chapter 3.
Enter the appropriate test values in the displayed
User Input screen. The input fields are described
below: The inputs required for this test are de-
scribed below.
n
S-Parameter: Drop down options. K Factor for
this measurement.
n
Data Points: Drop down options
- number of data
points in frequency sweep.
n
Frequency 1 (MHz): Lower frequency or start fre-
quency.
n
Frequency 2 (MHz): Upper frequency or stop fre-
quency.
n
Input Power (dBm):Input power level, in dBm.
n
IF Bandwidth (Hz):Intermediate frequency band-
width for test, in Hertz.
n
Test Set Setting: Byte value for setting test set
attenuation and source switch (below).
OPERATIONS, MEASUREMENT
S-PARAMETER TEST: K FACTOR
ME7840A OMM
5-7
Internal Source
Value
External Source
Value
0 dB
8
0 dB
0
10 dB
9
10 dB
1
20 dB
10
20 dB
2
30 dB
11
30 dB
3
40 dB
12
40 dB
4
50 dB
13
50 dB
5
60 dB
14
60 dB
6
70 dB
15
70 dB
7