•  Pressing the ON and OFF keys simultaneously will display 
the Version of software code. 
  
•  An audible beep will be made upon pressing of key pads 
and with the start and stop of bubble test. This beeping 
feature may be turned off by holding the ON key for five 
seconds. 
 
•  Each flow measurement is a true running average with the 
current test being added to the previous test and divided by 
the number of tests performed at that time. Average test 
number is displayed after each bubble test as A-1, A-2 etc. 
Pressing the ON key resets the averaging to start over and 
displays the current test.  
 
•  Upon turn-on, the display is “8888”  briefly, as a system 
check, and then becomes “0000”.  
 
•  The circuit board has advanced technology for improved 
accuracy and to maintain calibration with longer battery 
life. Up to 100 hours of usage on an over night charge. 
 
Original features of “EEEE” being displayed, if the reading has 
changed by ±5 % from previous reading or is out of the flow 
range remains in Version 2.   
 
 
IMPORTANT INFORMATION 
Your MSA DigiCal Calibrator has new 
 enhanced operational features.  
                     Effective October 1, 2005  Version 2  
FEATURES : 
Section 3 
 
Start Up Procedure 
 
A.   Read the previous section of Principle of Operation if unfamiliar with 
device. 
 
B.   Soap volume is not critical, therefore a measured amount is not 
necessary.  The proper amount is that which is sufficient to create a 
bubble.  Simply pour a small amount of soap through the bottom nipple 
to thoroughly cover the bottom of the flow cell and attach air source to 
nipple (bottom nipple for pressure, top nipple for vacuum).  Excessive 
amount of soap may cause a continuous stream of bubbles to go up the 
center tube at high flows. 
 
C.   Wet flow cell by connecting a pump flowing at 1000 to 2000 cc/min.   
Initiate soap film up center tube by rapidly pressing button down and 
releasing.  Repeat this procedure until bubble doesn't break.  The tube is 
now wet and tests at any range may be conducted.  One minute warm up 
time is all that is required of the DigiCal.