Error Code Memory
Display cycles through the last (3) 
Error Codes.
Erase by holding the Decrease (-) 
switches for 10 sec.
Currently have to hold (+) & (-)
 
LED DISPLAY 
 CALIBRATION MODE 
 DESCRIPTION
CALIBRATING THE DISPENSER
Pump Cal. Procedure
Set PRIME-NORMAL-RINSE switch 
to PRIME position. Place Graduated 
Cylinder under dispense tip. Press and 
hold Dispense Switch until pump stops 
(20 sec) Refer to Calibration Procedure.
NOTE: The LCA-1 is calibrated at the factory and does not normally need to be re-calibrated. Open the dispenser 
door to access the digital Programming Switches with LED display. To enter the Calibration Mode, Set the RUN/
NIGHT/PROGRAM Switch to the PROGRAM position and then hold both Up and Down Arrow switches until “1 
r u n” appears of the display, (approx 15 sec).
Pump Cal. #
Enter mL collected from the 20 sec. 
test above.
Default = 22
Range = 18 - 26mL
Water Flow Rate
Set PRIME-NORMAL-RINSE switch 
to RINSE position. Place Graduated 
Cylinder under dispense tip. Press and 
hold Dispense Switch until pump stops 
(20 sec) Refer to Calibration Procedure.
Water Flow Rate
Enter mL collected from the 20 sec. 
test above.
Default = 760
Range = 780 - 820 mLiters
Not used in the LCA-1
English or Metric Mode
Used to select the desired units.
Default = English
S.I. for Metric
MENU
(-)
(+)
1 r u n
MENU
(-)
(+)
1
2 2
MENU
(-)
(+)
2 r u n
MENU
(-)
(+)
2
6 0
MENU
(-)
(+)
3
MENU
(-)
(+)
4 E n g
MENU
(-)
(+)
5 E _ _