Chapter 9. Diagnostics and Troubleshooting
206
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
9.2 Displaying Diagnostic Parameters
As part of its measurement menu, the PT878 offers a list of diagnostic
parameters to aid in troubleshooting in the event of flowcell, transducer, or
electrical problems. You can select any diagnostic parameter for display as a
measurement as discussed in Chapter 5, Displaying and Configuring Data,
on page 95. Table 8 below shows all the available diagnostic parameters
and ranges.
parameters to aid in troubleshooting in the event of flowcell, transducer, or
electrical problems. You can select any diagnostic parameter for display as a
measurement as discussed in Chapter 5, Displaying and Configuring Data,
on page 95. Table 8 below shows all the available diagnostic parameters
and ranges.
Table 8: Diagnostic Parameters
Diagnostic
Parameter
Displays
Good
Bad
DT
Displays Delta T, or the difference between
the upstream and downstream transit
times.
the upstream and downstream transit
times.
N/A
Continuous
large fluctu-
ations of 1
ms or more.
Amplitude
Up
Displays the value for the amplitude dis-
criminator of the upstream transducer.
criminator of the upstream transducer.
20-28
<20 or >28
Amplitude
Down
Displays the value for the amplitude dis-
criminator of the downstream transducer.
criminator of the downstream transducer.
20-28
fluctuations
< 20 or > 28
fluctuations
fluctuations
T Up
Displays the upstream transit time of the
ultrasonic signal in micro seconds.
ultrasonic signal in micro seconds.
N/A
N/A
T Down
Displays the downstream transit time of
the ultrasonic signal in micro seconds.
the ultrasonic signal in micro seconds.
N/A
N/A
Gain/Up
Displays upstream gain in dB.
N/A
N/A
Gain/Dn
Displays downstream gain in dB.
N/A
N/A
Soundspeed
(m/s or f/s)
Displays the measured soundspeed of the
fluid.
fluid.
Check the soundspeed of
fluid in Soundspeeds and
Pipe Size Data.
P#up
Displays signal peaks for the upstream
transducer.
transducer.
100-900
<100, >900
P#dn
Displays signal peaks for the downstream
transducer.
transducer.
100-900
<100, >900
Reynolds # Displays the Reynolds Number.
N/A
N/A
K(RE)
Displays the K Factor, based on the
Reynolds Number.
Reynolds Number.
N/A
N/A
Raw Up
Displays raw upstream signal.
N/A
N/A
Raw Down
Displays raw downstream signal.
N/A
N/A