Chapter 9. Diagnostics and Troubleshooting
206
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
9.2 Displaying Diagnostic Parameters
As part of its measurement menu, the PT878 offers a list of diagnostic 
parameters to aid in troubleshooting in the event of flowcell, transducer, or 
electrical problems. You can select any diagnostic parameter for display as a 
measurement as discussed in Chapter 5, Displaying and Configuring Data, 
on page 95. Table 8 below shows all the available diagnostic parameters 
and ranges.
Table 8:  Diagnostic Parameters
Diagnostic
Parameter
 Displays
Good
Bad
DT
Displays Delta T, or the difference between 
the upstream and downstream transit 
times.
N/A
Continuous 
large fluctu-
ations of 1 
ms or more.
Amplitude 
Up
Displays the value for the amplitude dis-
criminator of the upstream transducer.
20-28
<20 or >28
Amplitude 
Down
Displays the value for the amplitude dis-
criminator of the downstream transducer.
20-28 
fluctuations
< 20 or > 28 
fluctuations
T Up
Displays the upstream transit time of the 
ultrasonic signal in micro seconds.
N/A
N/A
T Down
Displays the downstream transit time of 
the ultrasonic signal in micro seconds.
N/A
N/A
Gain/Up
Displays upstream gain in dB.
N/A
N/A
Gain/Dn
Displays downstream gain in dB.
N/A
N/A
Soundspeed
(m/s or f/s)
Displays the measured soundspeed of the 
fluid.
Check the soundspeed of 
fluid in Soundspeeds and 
Pipe Size Data.
P#up
Displays signal peaks for the upstream 
transducer.
100-900
<100, >900
P#dn
Displays signal peaks for the downstream 
transducer.
100-900
 <100, >900
Reynolds # Displays the Reynolds Number.
N/A
N/A
K(RE)
Displays the K Factor, based on the 
Reynolds Number.
N/A
N/A
Raw Up
Displays raw upstream signal.
N/A
N/A
Raw Down
Displays raw downstream signal.
N/A
N/A