Chapter 3. Programming Site Data
30
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
3.2.2 Confirming Entries
To confirm the entries and return to Operate mode, press 
[F3]
 (OK). 
To leave the window without confirming the entries, press 
[F2]
 (Cancel) 
or the 
[ESC]
 key.
In either case, the PT878 returns to Operate Mode.
3.3 Entering Pipe Parameters
To enter the Pipe option, scroll to the Pipe entry on the Program Menu and 
press 
[ENTER]
. (From the Transducer window, you can scroll back up to the 
Transducer tab and press the [
X
] arrow key to reach the Pipe window, and 
press 
[ENTER]
.) The screen appears similar to Figure 15 below. To step 
through each parameter, press the [
T
of Appendix A, Menu Maps.
Note:
Refer to the brochure Soundspeeds and Pipe Size Data (914-004) for 
additional information about pipe sizes and soundspeeds.
Figure 15: Pipe Option Window
Transducer
Lining
Path
Transducer/Pipe
Pipe
Fluid
Cancel
OK
Material
Other
Sound Speed 
600.3 m/s
5
ANSI
OD, mm
Wall, mm
2
Nominal       Schedule
OD x PI, mm
15.708
Measure Wall with TGauge