  7) Pump does not circulate the water. 
   
• Debris may be in the pump impeller housing. Raise the   
      shroud unit and lock in the up position. Inspect pump for 
      debris and remove with tweezers or similar instrument. 
 
  8) HEATING light is flashing. 
   
• This is normal operation. At approximately one degree   
 
below the set temperature, the microprocessor begins to   
 
reduce the amount of heat generated and the HEATING 
 
light begins to flash.   
 
  9) Temperature display at WATER TEMP deviates from 
   thermometer reading. 
   
• Run the unit for more than 30 minutes. 
   
• Read thermometer temperature while the thermometer 
      remains in water. Remember the temperature reading. 
   
• Press the POWER key to turn the unit power off. 
   
• Press the SET TEMP key and the POWER key at the 
      same time. Power is turned on again but the WATER 
      TEMP display shows an arbitrary number.   
   
• Each time the [5] key is pressed, the number displayed 
      in WATER TEMP window increases by 0.1. Each time 
      the [0] key is pressed, it decreases by 0.1. 
      Press either the [5] key or the [0] key until the WATER 
      TEMP shows same number as the thermometer reading. 
      This adjustment should be done in one minute from 
      when you have read the thermometer. 
   
• Press the POWER key to turn power off. 
   
• Press the POWER key to turn power on again and 
      check calibration has been done correctly. 
 
 
 