CALIBRATION MENU
14 
CAL TEMP
OK  
Å Æ
CAL RH%
OK  
Å Æ
RH CYCLE
OK  
Å Æ
AMBIENT
OK  
Å Æ
SAVE
OK  
Å Æ
CANCEL
OK  
Å Æ
T 37.3C
OK  CAL!
RH 28%
OK  CAL!
RHC 7S
OK  
-  +
T 20.0C
OK  
-  +
CALIBRATE THERMOMETER.
TAKE READINGS AT SEVERAL
PLACES 20mm ABOVE THE BASE OF
THE EMPTY EGG BASKET AND
CALCULATE THE AVERAGE.
CALIBRATE HYGROMETER.
PLACE HYGROMETER IN EGG
BASKET WITH NO WATER IN
TROUGHS.
HUMIDITY PUMP CYCLE TIME.
ONLY APPLICABLE WHEN USED
WITH THE BRINSEA ADVANCE
HUMIDITY PUMP.
AMBIENT TEMPERATURE
COMPENSATION.
AMBIENT TEMPERATURE USED IN
HEATER CONTROL ALGORITHM.
ONLY ADJUST IF DISPLAY VARIES
BY MORE THAN +/- 0.2°C.
PRESS ALL 3 BUTTONS TO UNLOCK THE CALIBRATION MENU
SELECT THE OPTION / RETURN TO THE MENU.
GO FORWARD ONE SCREEN / INCREASE THE VALUE.
GO BACK ONE SCREEN / DECREASE THE VALUE.