SITRANS LC 500 – INSTRUCTION MANUAL 
 7ML19985GE01
mmm
mm
Pr
obe Configur
at
io
n
SITRANS LC 500: Probe Configuration
The probe (electrode) comprises a measurement section and an active shield section. 
This electrode connects to the capacitance detector portion of the two-wire loop 
powered electronic transmitter. The transmitter module is mounted in a powder-coated 
aluminum enclosure which provides reliable operation in environments with dust, 
moisture, and high-frequency interference. 
SITRANS LC 500 Electrode (Probe) Characteristics
Apply to all general connection configurations:
The standard SITRANS LC 500 insulated electrode is designed for use in both 
conducting and non-conducting liquid applications. 
Most electrodes consist of an active shield portion and a measurement portion, 
which combine to form the complete electrode. (This is not the case for cable 
electrodes or electrodes with ceramic/enamel insulation.)
The sum of the active shield length and the measurement length is the total 
insertion length. 
The active shield design provides continuous immunity from changes in conditions 
at the top of the vessel where levels of vapors, dust, and condensation may be 
constantly changing.
The design of the active shield isolates the starting capacitance of the electrode 
from the effects of changes in capacitance due to temperature and pressure 
fluctuations that could cause small changes in the seal geometry. 
The carefully-controlled diameter of the electrodes and insulation produces a linear 
output over a wide range of capacitance values (1 pF to 3300 pF). 
The end seal is formed as an integral part of the electrode insulation, giving smooth 
and uniform characteristics (tested to 55 kV).
Standard single cone seal 
insertion length
active 
shield
measure-
ment 
portion
probe 
seal 
(inactive)