SITRANS LC 500 – INSTRUCTION MANUAL 
Page 45
mm
mmm
St
art
 Up
Select capacitance or percent mode for dynamic primary variable (PV): 
menu 01
a. To see the value displayed as percent, select menu 01 and press both buttons to 
set Pv = P.
b. To see the value displayed as pF, select menu 01 and increase or decrease the 
value till Pv = 0.
The SITRANS LC 500 is now ready to operate. For a table showing the different functions 
available, and the combinations of switch position and button presses used to carry out 
these functions, see page 41. For a detailed list of menu items, see 
Notes: 
During normal operation, the 4 and/or 20 mA point can be calibrated at any time.
If the difference in the capacitance value between the 4 mA point and the 20 mA 
point is smaller than the minimum span value (3.3 pF), the new value will not be 
accepted.
To revert to factory settings, select menu 12 and press both buttons: the display 
will read FAC A.