SITRANS LC 500 – INSTRUCTION MANUAL 
7ML19985GE01
mmmm
m
Appendix C
MultiDrop operation
The SITRANS LC 500 transmitter supports MultiDrop Operation.
Burst mode
The SITRANS LC 500 transmitter does not support Burst Mode.
Units conversions
The Transmitter Variable #0 Units are in pF and cannot be changed. 
The Transmitter Variable #1 Values may be set to any Units and Value with Command #140. 
The Transmitter Variable Range Values may be read at any time with Command #141. 
The value returned as Secondary Variable (S.V.) is the result of the following calculation:
TV1 = TV0 Dynamic Range Value in percent x ({TV#1}URV – {TV#1}LRV) + {TV#1}LRV.
This method provides a means of converting TV#0, which is always in pF, to alternative 
units (level or volume).
Additional Universal Command Specifications
For a document listing the additional Universal Command Specifications, please contact 
Technical Publications at techpubs@siemens-milltronics.com