 DEVICE DESCRIPTION 
2
13
SITRANS FX300
www.siemens.com/flow
06/2009 - SFIDK.PS.050.F2.02   A5E02100423
2.2.3  Device description
2.2.4  Free air delivery measurement - FAD (optional)
A (air) compressor draws air from the ambient atmosphere, compresses it and delivers it at the 
required pressure. Since the ambient atmosphere also contains water vapour, what the 
compressor draws in is a mixture of air and water vapour. Free air delivery measurement is to 
be understood under this condition. Most manufacturers specify free air delivery only at 
standard intake conditions. What the user ultimately requires as process air must first be 
determined before measuring can take place with an accuracy of ±1%.
The vortex flowmeter with the optional FAD function can measure the free air delivery online, 
humidity and speed compensated, regardless of its function as standard flowmeter. The 
integrated software evaluates the free air delivery automatically online.
The menu-driven, user-friendly software prompts the operator to enter the pressure, relative 
humidity, the required as well as current discharge pressure.
The steam tables and compressibility tables are saved as standard. The measuring device is 
optionally available with a pressure sensor which measures the discharge pressure online, 
making manual input of the values unnecessary.
Figure 2-4: Device description
1  Signal converter
2  Cable feedthrough grey, standard version
3  Pressure sensor, optional
4  Shut-off valve, optional
5  Measuring sensor
6  Centering ring
MA_Siemens_Sitrans_FX300_R02_en_PRT.book  Page 13  Monday, June 22, 2009  8:17 AM