 OPERATION 
6
47
SITRANS FX300
www.siemens.com/flow
06/2009 - SFIDK.PS.050.F2.02   A5E02100423
6.5  Menu structure
6.5.1  Overview of firmware versions
There are three firmware versions, each of which is tailored to a different use of the 
measuring device:
• Basic:
Basic:
Basic:
Basic: liquids and gases without compensation, saturated steam with density compensation 
using the temperature
• Steam:
Steam:
Steam:
Steam: saturated and superheated steam with density compensation using pressure and 
temperature, gross heat meter
• Gas:
Gas:
Gas:
Gas: gas, gas mixture and wet gas with density compensation using pressure and 
temperature, FAD (free air delivery measurement)
The menu structures differ from one another depending on the firmware version.
The following table contains an overview of all menu items in the first menu level. For a 
complete description of the menu, note the firmware version of your measuring device and 
follow the relevant instructions in the table.
Menu items
Firmware version
Basic
Firmware version
Steam
Firmware version
Gas
1. Quick setup
 refer to 
Menu item Quick Setup
 on page 49
2. Tests
 refer to 
Menu item Tests
 on page 50
3. Setup
 refer to 
Menu item 
Setup (firmware version 
- basic)
 on page 51
 refer to 
Menu item 
Setup (firmware version 
- steam)
 on page 54
 refer to 
Menu item 
Setup (firmware version 
- gas)
 on page 58
5. Service
The service menu is only accessible to service personnel and is not described.
Table 6-9: Overview of menu structure
INFORMATION!
The measuring device has been preset at the factory in accordance with the customer order. 
Therefore, subsequent configuration via the menu is only necessary if the intended use of the 
measuring device changes.