Philips Vortex Flow Meter F-2200 Series Manual De Usuario

Descargar
Página de 22
F-2200 Series Vortex Flow Meter Installation and Operation Guide • Revised 12/08
12 
The diagrams shown below illustrate the correct meter orientation for vertical and horizontal pipe. 
Nominal Meter Size 
(in.)
Meter Body 
Diameter (in.)
Maximum Pipe ID 
Difference (in.)
1”
1.05
0.016
1 1/2”
1.61
0.016
2”
2.07
0.024
3”
3.07
0.024
4”
4.03
0.024
6”
6.07
0.031
8”
7.98
0.039
 
 
Ensure that the bore of the locating pipes are smooth and without deposits, 
 
 
scaling, or welding beads. 
 
3.1.2  Maximum Allowable Difference Between Inside Diameter Of Meter Body 
        
And The Connecting Pipe 
 
 
The table below provides the maximum allowable difference between the diameter of 
   the 
fl ow meter body and the connecting pipe.