Communications Setup and Operation 
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
104  
 
Enabled mode (32) are selected, the analyzer would display a combined MODE 
ID
 of 35.   
Table 6-1:  COM Port Communication Modes 
MODE
1
 ID 
DESCRIPTION 
QUIET 
Quiet mode suppresses any feedback from the analyzer (such as warning messages) to 
the remote device and is typically used when the port is communicating with a computer 
program where such intermittent messages might cause communication problems.   
Such feedback is still available but a command must be issued to receive them. 
COMPUTER  
Computer mode inhibits echoing of typed characters and is used when the port is 
communicating with a computer operated control program. 
HESSEN 
PROTOCOL
 
16
 
The Hessen communications protocol is used in some European countries. TELEDYNE 
API PN 02252 contains more information on this protocol. 
E, 8, 1 
8192 
When turned on this mode switches the COM port settings from 
● 
NO PARITY; 8 data bits; 1 stop bit  to  EVEN PARITY; 8 data bits; 1 stop bit.
 
E, 7, 1 
2048 
When turned on this mode switches the COM port settings from 
● 
NO PARITY;  8 DATA BITS; 1 stop bit  to  EVEN PARITY;  7 DATA BITS; 1 stop bit.
 
RS-485 
1024 
Configures the COM2 Port for RS-485 communication.  RS-485 mode has precedence 
over multidrop mode if both are enabled. 
SECURITY  
When enabled, the serial port requires a password before it will respond (see Section  
8.2.1.6).  The only command that is active is the help screen (?  CR).   
MULTIDROP 
PROTOCOL 
32 
Multidrop protocol allows a multi-instrument configuration on a single communications 
channel.  Multidrop requires the use of instrument IDs. 
ENABLE 
MODEM 
64 
Enables to send a modem initialization string at power-up.  Asserts certain lines in the 
RS-232 port to enable the modem to communicate. 
ERROR 
CHECKING
128 
Fixes certain types of parity errors at certain Hessen protocol installations. 
XON/XOFF 
HANDSHAKE
2
 
256 
Disables XON/XOFF data flow control also known as software handshaking. 
HARDWARE 
HANDSHAKE 
Enables CTS/RTS style hardwired transmission handshaking.  This style of data 
transmission handshaking is commonly used with modems or terminal emulation 
protocols as well as by Teledyne Instrument’s APICOM software. 
HARDWARE 
FIFO 
512 
Disables the HARDWARE FIFO (First In – First Out), When FIFO is enabled it improves 
data transfer rate for that COM port. 
COMMAND 
PROMPT 
4096 
Enables a command prompt when in terminal mode. 
1
 Modes are listed in the order in which they appear in the  
SETUP  MORE  COM  COM[1 OR 2]  MODE menu 
2
 The default setting for this feature is ON.  Do not disable unless instructed to by Teledyne API Technical Support 
personnel. 
 
Note 
Communication Modes for each COM port must be configured 
independently. 
 