Teledyne Tablet Accessory T803 Manual De Usuario
Communications Setup and Operation
Teledyne API T803 CO2/O2 Analyzer Operation Manual
116
5. In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following
settings:
Baud Rate: 115200
COM2 Mode Settings:
Quiet Mode
ON
SECURITY MODE
OFF
Computer Mode
ON
MULTIDROP MODE
OFF
MODBUS RTU
OFF
ENABLE MODEM
OFF
MODBUS ASCII
OFF
ERROR CHECKING
ON
E,8,1 MODE
OFF
XON/XOFF HANDSHAKE
OFF
E,7,1 MODE
OFF
HARDWARE HANDSHAKE OFF
RS-485 MODE
OFF
HARDWARE FIFO
ON
COMMAND PROMPT
OFF
6. Next, configure your communications software, such as APIcom. Use the
COM port determined in Step 4 and the baud rate set in Step 5. The figures
below show how these parameters would be configured in the Instrument
Properties window in APIcom when configuring a new instrument. See the
APIcom manual (PN 07463) for more details.
below show how these parameters would be configured in the Instrument
Properties window in APIcom when configuring a new instrument. See the
APIcom manual (PN 07463) for more details.
Note
USB configuration requires that the baud rates of the instrument
and the PC match; check the PC baud rate and change if needed.
and the PC match; check the PC baud rate and change if needed.
Using the USB port disallows use of the rear panel COM2 port
except for multidrop communication.
except for multidrop communication.