Communications Setup and Operation 
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
116  
 
5.  In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following 
settings: 
Baud Rate:  115200 
 
 
 
COM2 Mode Settings: 
 
 
 
Quiet Mode 
ON 
SECURITY MODE 
OFF
Computer Mode 
ON 
MULTIDROP MODE 
OFF
MODBUS RTU 
OFF
ENABLE MODEM 
OFF
MODBUS ASCII 
OFF
ERROR CHECKING 
ON 
E,8,1 MODE 
OFF
XON/XOFF HANDSHAKE 
OFF
E,7,1 MODE 
OFF
HARDWARE HANDSHAKE   OFF
RS-485 MODE 
OFF
HARDWARE FIFO 
ON 
 
 
COMMAND PROMPT 
OFF
6.  Next, configure your communications software, such as APIcom.  Use the 
COM port determined in Step 4 and the baud rate set in Step 5.   The figures 
below show how these parameters would be configured in the Instrument 
Properties window in APIcom when configuring a new instrument.  See the 
APIcom manual (PN 07463) for more details. 
 
 
Note 
 
USB configuration requires that the baud rates of the instrument 
and the PC match; check the PC baud rate and change if needed. 
 
Using the USB port disallows use of the rear panel COM2 port 
except for multidrop communication.  