Teledyne API T803 CO2/O2 Analyzer Operation Manual 
Data Acquisition System (DAS & APICOM 
 
 131 
7.1.2.1. DAS CHANNELS 
The key to the flexibility of the DAS is its ability to store a large number of 
combinations of triggering events and data parameters in the form of data 
channels.  Users may create up to 50 data channels and each channel can contain 
one or more parameters.  For each channel, the following are selected: 
  One triggering event is selected 
  Up to 50 data parameters, which can be the shared between channels.   
  Several other properties that define the structure of the channel and allow the 
user to make operational decisions regarding the channel. 
Table 7-2:  DAS Data Channel Properties 
PROPERTY DESCRIPTION 
DEFAULT 
SETTING 
SETTING RANGE 
NAME 
The name of the data channel. 
“NONE” 
Up to 6 letters or digits 
1.  
TRIGGERING 
EVENT 
The event that triggers the data channel to measure 
and store the datum 
ATIMER 
Any available event  
(see Appendix A-5). 
NUMBER AND 
LIST OF 
PARAMETERS 
A User-configurable list of data types to be 
recorded in any given channel. 
(PMTDET) 
Any available parameter 
(see Appendix A-5). 
REPORT PERIOD 
The amount of time between each channel data 
point. 
000:01:00 
(1 hour)
 
000:00:01 to 
366:23:59 
(Days:Hours:Minutes) 
NUMBER OF 
RECORDS 
The number of reports that will be stored in the data 
file.  Once the limit is exceeded, the oldest data is 
over-written. 
100 
Configuration-dependent, 
limited by available 
storage space. 
RS-232 REPORT 
Enables the analyzer to automatically report 
channel values to the RS-232 ports. 
OFF 
OFF or ON 
CHANNEL 
ENABLED 
Enables or disables the channel.  Allows a channel 
to be temporarily turned off without deleting it. 
ON 
OFF or ON 
CAL HOLD OFF 
Disables sampling of data parameters while 
instrument is in calibration mode 
2
OFF 
OFF or ON 
1
 More with APICOM, but only the first six are displayed on the front panel). 
2
 When enabled records are not recorded until the DAS HOLDOFF period is passed after calibration mode.  DAS HOLDOFF SET in the 
VARS menu (see Section
.) 
  
7.1.3. DEFAULT DAS CHANNELS 
CONC: Samples CO
2
 and O
2
 concentrations at one-minute intervals and stores an 
average every five minutes with a time and date stamp.  Readings during 
calibration and calibration HOLD OFF are not included in the data.. 
PNUMTC: Collects sample flow and sample pressure data at five-minute 
intervals and stores an average once a day with a time and date stamp.  This data 
is useful for monitoring the condition of the pump and critical flow orifice 
(sample flow) and the sample filter (clogging indicated by a drop in sample 
pressure) over time to predict when maintenance will be required.  The last 360 
daily averages (about 1 year) are stored. 