Troubleshooting and Service 
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
196 
 
remote commands and the display changes accordingly, the touchscreen interface 
may be faulty. 
11.6.5. LCD DISPLAY MODULE 
Verify the functioning of the front panel display by observing it when power is 
applied to the instrument. Assuming that there are no wiring problems and that 
the DC power supplies are operating properly, the display screen should light and 
show the splash screen with logo and other indications of its state as the CPU 
goes through its initialization process. 
11.6.6. RELAY BOARD 
The relay board PCA (04135) can be most easily checked by observing the 
condition of the its status LEDs on the relay board, as described in Section 
12.6.3.1, and the associated output when toggled on and off through signal I/O 
function in the diagnostic menu, See Section 11.1.3.   
1.  If the front panel display responds to key presses and D1 on the relay board 
is NOT flashing then either the wiring between the Keyboard and the relay 
board is bad, or the relay board is bad. 
2.  If D1 on the relay board is flashing and the status indicator for the output in 
question (heater power etc.) toggles properly using the signal I/O function, 
then the associated control device on the relay board is bad.  (Several of the 
control devices are in sockets and can be easily replaced).   
11.6.7. SENSOR ASSEMBLY 
 Both the CO
and O
2
 sensors have no user serviceable parts. 
11.6.8. PRESSURE/FLOW SENSOR ASSEMBLY 
The pressure/flow sensor PCA can be checked with a Voltmeter using the 
following procedure which, assumes that the wiring is intact, and that the 
motherboard and the power supplies are operating properly: 
07276B DCN6418