Teledyne API T803 CO2/O2 Analyzer Operation Manual 
Troubleshooting and Service 
 
 
199 
11.6.9.3. STATUS OUTPUTS 
The procedure below can be used to test the Status outputs: 
1.  Connect a jumper between the “D“ pin and the “” pin on the status output 
connector. 
2.  Connect a 1000 ohm resistor between the “+” pin and the pin for the status 
output that is being tested. 
3.  Connect a voltmeter between the “” pin and the pin of the output being 
tested (see table below).   
Under the DIAG SIGNAL I/O menu (See Section11.1.3), scroll through the 
inputs and outputs until you get to the output in question.  Alternately turn on and 
off the output noting the voltage on the voltmeter, it should vary between 0 volts 
for ON and 5 volts for OFF.   
Table 11-8:  Status Outputs Check 
PIN (LEFT TO RIGHT) 
STATUS 
SYSTEM OK / ALARM 
CONC VALID / CONC 
WARNING 
 CAILIBRATION MODE / 
MEASURE MODE 
4 SPAN 
/ZERO 
CAL 
RNG2 / RNG1 CAL 
6 CO
/ O
 SENSOR  CAL 
7 SPARE 
8 SPARE 
 
11.6.10. CPU 
There are two major types of failures associated with the CPU board: complete 
failure and a failure associated with the Disk-On-Module (DOM) on the CPU 
board.  If either of these failures occurs, contact the factory.   
For complete failures, assuming that the power supplies are operating properly 
and the wiring is intact, the CPU is faulty if on power-on, the watchdog LED on 
the motherboard is not flashing. 
In some rare circumstances, this failure may be caused by a bad IC on the 
motherboard, specifically U57, the large, 44 pin device on the lower right hand 
side of the board. If this is true, removing U57 from its socket will allow the 
instrument to start up but the measurements will be invalid. 
If the analyzer stops during initialization (the front panel display shows a fault or 
warning message), it is likely that the DOM, the firmware or the configuration 
and data files have been corrupted.  
 