Teledyne API T803 CO2/O2 Analyzer Operation Manual 
Principles of Operation 
 
211 
SPRING 
O-RINGS 
FILTER 
CRITICAL 
FLOW 
ORIFICE 
AREA OF 
LOW 
PRESSURE 
AREA OF 
HIGH 
PRESSURE 
Sonic 
Shockwave 
 
Figure 12-6: 
Flow Control Assembly & Critical Flow Orifice 
The actual flow rate of gas through the orifice (volume of gas per unit of time), 
depends on the size and shape of the aperture in the orifice.  The larger the hole, 
the more gas molecules, moving at the speed of sound, pass through the orifice.  
Because the flow rate of gas through the orifice is only related to the minimum 
2:1 pressure differential and not absolute pressure the flow rate of the gas is also 
unaffected by degradations in pump efficiency due to age.   
The critical flow orifice used in the T803 is designed to provide a flow rate of 
120 cm
3
/min. 
12.5.2. PARTICULATE FILTER  
The T803 Analyzer comes equipped with a 47 mm diameter, Teflon, particulate 
filter with a 1 micron pore size.  The filter is accessible through the front panel, 
which folds down to allow access, and should be changed according to the 
suggested maintenance schedule described in  
12.5.3. PNEUMATIC SENSORS 
12.5.3.1. SAMPLE PRESSURE SENSOR 
An absolute value pressure transducer plumbed to the outlet of the sample 
chamber is used to measure sample pressure.  The output of the sensor is used to 
compensate the concentration measurement for changes in ambient air pressure.  
This sensor is mounted to a printed circuit board with the sample flow sensor on 
the sample chamber; see the following section and Figure 3-5. 
12.5.3.2. SAMPLE FLOW SENSOR  
A thermal-mass flow sensor is used to measure the sample flow through the 
analyzer.  The sensor is calibrated at the factory with ambient air or N
2
, but can 
be calibrated to operate with samples consisting of other gases such as O
2
.  This 
07276B DCN6418