Principles of Operation 
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
220 
 
12.6.7. POWER SUPPLY / CIRCUIT BREAKER 
The analyzer operates on 100 VAC, 115 VAC or 230 VAC power at either 50Hz 
or 60Hz.  Individual units are set up at the factory to accept any combination of 
these five attributes.  As illustrated in Figure 12-12, power enters the analyzer 
through a standard IEC 320 power receptacle located on the rear panel of the 
instrument.  From there it is routed through the On/Off switch located in the 
lower right corner of the Front Panel.  A 6.75 Amp circuit breaker is built into the 
ON/OFF Switch. 
AC power is distributed directly to the sample gas pump. 
AC Line power is stepped down and converted to DC power by two DC power 
supplies.  One supplies +12 VDC, while a second supply provides +5 VDC and 
±15 VDC for logic and analog circuitry.  All DC voltages are distributed via the 
relay board. 
 
 
CAUTION  
G
ENERAL 
S
AFETY 
H
AZARD
 
Should the AC power circuit breaker trip, investigate and correct the condition causing 
this situation before turning the analyzer back on. 
 
 
Figure 12-12:  Power Distribution Block Diagram 
07276B DCN6418