Getting Started 
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
24    
This record, Final Test and Validation Data Sheet, PN 068360000
 
is an 
important quality assurance and calibration record for this instrument.  It 
should be placed in the quality records file for this instrument. 
1.  Carefully remove the top cover of the analyzer and check for internal shipping 
damage as follows: 
a)  Remove the locking screw located in the top, center of the Front panel; 
b)  Remove the two flat head, Phillips screws on the sides of the instrument 
(one per side towards the rear); 
c)  Slide the cover backwards until it clears the analyzer’s front bezel;  
d)  Lift the cover straight up. 
2.  Inspect the interior of the instrument to make sure all circuit boards and other 
components are in good shape and properly seated.   
3.  Check that the connectors of the various internal wiring harnesses and 
pneumatic hoses are firmly and properly seated. 
4.  Verify that all of the optional hardware ordered with the unit has been 
installed.  These are listed on the paperwork accompanying the analyzer. 
 
 
 