Teledyne API T803 CO2/O2 Analyzer Operation Manual 
Getting Started 
 
 33 
 
CAUTION
 
G
ENERAL 
S
AFETY 
H
AZARD
 
The T803 analyzer can be configured for both 100-130 V and 210-240 V at either 
47 or 63 Hz.   
To avoid damage to your analyzer, make sure that the AC power voltage 
matches the voltage indicated on the Analyzer’s serial number label tag (See 
Figure 3-4) b
efore plugging the T803 into line power.
 
 
3.3.1.2. ANALOG INPUTS (OPTION 64) CONNECTIONS 
The Analog In connector is used for connecting external voltage signals from 
other instrumentation (such as meteorological instruments) and for logging these 
signals in the analyzer’s internal DAS. The input voltage range for each analog 
input is 0-10 VDC, and the input impedance is nominally 20kΩ in parallel with 
0.1µF. 
 
Figure 3-6: 
Analog In Connector 
 
Pin assignments for the Analog In connector are presented in Table 3-4. 
Table 3-4.  Analog Input Pin Assignments 
PIN DESCRIPTION 
DAS 
PARAMETER
1
 
Analog input # 1 
AIN 1 
Analog input # 2 
AIN 2 
Analog input # 3 
AIN 3 
Analog input # 4 
AIN 4 
Analog input # 5 
AIN 5 
Analog input # 6 
AIN 6 
Analog input # 7 
AIN 7 
Analog input # 8 
AIN 8 
GND 
Analog input Ground 
N/A 
1
 See Section 7.1 for details on setting up the DAS. 
07276B DCN6418