Teledyne API T803 CO2/O2 Analyzer Operation Manual 
Setup Menu 
 
 71 
 
(The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 
5.4.4. SETUP RNGE  DIL: USING THE OPTIONAL DILUTION RATIO 
FEATURE 
This feature is an optional software utility that is used to compensate for any 
dilution of the sample gas that may occur before it enters the sample inlet. 
Typically this occurs in continuous emission monitoring (CEM) applications 
where the sampling method used to remove the gas from the stack, dilutes the 
sample.   
Using the dilution ratio option is a 3-step process: 
1.  Select the reporting range mode and set the reporting range upper limit (see 
Section 5.4.3).   
Ensure that the upper span limit entered for the reporting range is the 
maximum expected concentration of the non-diluted gas.  