Setup Menu 
 
 93 
current meter (see Figure 3-7 for pin assignments and diagram of the analog 
output connector).  This allows the use of a voltmeter connected across the 
resistor to measure converter output as VDC or mVDC. 
+DC     Gnd    
 
Figure 5-4: 
Alternative Setup Using 250Ω Resistor for Checking Current Output Signal Levels 
 
In this case, follow the procedure above but adjust the output for the following 
values: 
 
Table 5-6: 
Current Loop Output Check 
% FS 
Voltage across 
Resistor for 2-20 mA 
Voltage across 
Resistor for  4-20 mA 
500 mVDC 
1000 mVDC 
100 
5000 mVDC 
5000 mVDC 
 
07276B DCN6418