Teledyne Tablet Accessory T803 Manual De Usuario
Teledyne API T803 CO2/O2 Analyzer Operation Manual
Setup Menu
93
current meter (see Figure 3-7 for pin assignments and diagram of the analog
output connector). This allows the use of a voltmeter connected across the
resistor to measure converter output as VDC or mVDC.
output connector). This allows the use of a voltmeter connected across the
resistor to measure converter output as VDC or mVDC.
V
+DC Gnd
Figure 5-4:
Alternative Setup Using 250Ω Resistor for Checking Current Output Signal Levels
In this case, follow the procedure above but adjust the output for the following
values:
values:
Table 5-6:
Current Loop Output Check
% FS
Voltage across
Resistor for 2-20 mA
Voltage across
Resistor for 4-20 mA
0
500 mVDC
1000 mVDC
100
5000 mVDC
5000 mVDC
07276B DCN6418