13.1.3. UV SOURCE LAMP 
The source of excitation UV light for the T100 is a low pressure zinc-vapor lamp. An 
AC voltage heats up and vaporizes zinc contained in the lamp element creating a light-
producing plasma arc. Zinc-vapor lamps are preferred over the more common mercury-
vapor lamps for this application because they produce very strong emission levels at the 
wavelength required to convert SO
2
 to SO
2
The lamp used in the T100 is constructed with a vacuum jacket surrounding a double-
bore lamp element (refer to Figure 13-3). The vacuum jacket isolates the plasma arc 
from most external temperature fluctuations. The jacket also contains thermal energy 
created by the lamp’s operation therefore helping the lamp to heat up and maintain 
proper vaporization temperature. Light is emitted through a 20 mm x 5 mm portal.  
06807C DCN6650