National Instruments Welding System 321645c-01 Manual De Usuario

Descargar
Página de 554
Chapter 2
Function Reference — GPCTR_Set_Application
©
 National Instruments Corporation
2-239
NI-DAQ FRM for PC Compatibles
Figure 2-26.  Buffered Period Measurement when No Source Edges Are Present between Gate Edges
application = 
ND_BUFFERED_SEMI_PERIOD_MSR
In this application, the counter is used for the continuous measurement of the time interval 
between successive transitions of the gate signal. By default, those are all transitions of the 
signal on the line given in Table 2-25. The counter counts the 20 MHz internal timebase 
(
ND_INTERNAL_20_MHZ
), so the resolution of measurement is 50 ns. The counter counts up 
starting from 0; its contents are placed in the buffer after an edge is detected on the gate; the 
counter then starts counting up from 0 again. NI-DAQ transfers data from the counter into the 
buffer until the buffer is filled; the counter is disarmed at that time.
Note
The counter will start counting as soon as you arm it. Be aware of this when you 
interpret the first count in your buffer.
Figure 2-27.  Buffered Semi-Period Measurement when No Source Edges Are Present between Gate Edges
Source
Gate
Buffer
2
1
4
3
6
5
1
7
8
9
3
2
5
4
1
6
3
2
4
Measured
Period
Measured
Period
Measured
Period
Measured
Period
9
9
9
9
9
9
9
9
 9
 4
The instant
you arm
the counter
Warning
Source
Buffer
3
Gate
1
3
1
2
4
5
2
4
1
2
1
2
3
1
2
1
2
Measured
Semi-Period
3
Measured
Semi-Period
Measured
Semi-Period
Measured
Semi-Period
3
3
 5
3
2
 4
3
2
2
 4
3
2
2
5
 4
3
2
2
2
5
 4
3
2